云南燕子岩观测站DSQ水管倾斜仪潮汐扰动现象研究

Research on tidal disturbance phenomenon of DSQ water tube tiltmeter at Yunnan Yanziyan observation station

  • 摘要: 云南燕子岩观测站DSQ水管倾斜仪各分量频繁记录到高频扰动,表现形式为观测曲线脉动加粗,更换放大盒、线圈、线路、数采等都未能排除此高频扰动。经与实况气象观测资料对比后发现:潮汐扰动与气压波动有一定的同步性。经进一步诊断、排查、分析,确定引起高频扰动的原因为:DSQ水管倾斜仪钵体罩子缺少空气管,导致水管仪两端液面上方各自形成独立空腔,两端液面压力平衡被打破,引起高频扰动。对钵体罩子进行打孔处理后,高频扰动排除。这种高频扰动诊断与排除的过程和方法,对提高DSQ水管倾斜仪固体潮观测质量具有重要意义,对改进、管理同类仪器具有参考和借鉴价值。

     

    Abstract: The DSQ water tube tiltmeter at Yunnan Yanziyan observation station frequently records high-frequency disturbances in various components, manifested in the form of thickened observation curve pulsation, and the replacement of amplification boxes, coils, lines, data acquisition, etc. failed to eliminate this high-frequency disturbance. After comparing with actual meteorological observation data, it was found that there was a certain degree of synchronization between tidal disturbances and atmospheric pressure fluctuations. After further diagnosis, investigation, and analysis, it had been determined that the cause of high-frequency disturbance was that the bowl cover of the DSQ water tube tiltmeter lacked air pipes, resulting in independent cavities above the liquid levels at both ends of the tiltmeter. The pressure balance at both ends of the liquid level was disrupted, causing high-frequency disturbances. After punching the bowl cover, high-frequency disturbances were eliminated. The process and method of diagnosing and eliminating high-frequency disturbances are of great significance for improving the quality of solid tide observation using the DSQ water tube tiltmeter, and have reference value for improving and managing similar instruments.

     

/